Tag

Jaszczak phantom
 (

A Jaszczak phantom () aka Data Spectrum ECT phantom Quality assurance for SPECT systems. IAEA Health Series No. 6. 2009. International Atomic Energy Agency publications. p.182 is an used for validating scanner geometry, 3D contrast, uniformity, resolution, attenuation and scatter correction or alignment tasks in . It is commonly used in academic centers and hospitals to characterize a or some gamma camera systems for purposes. It is used for accreditation by clinical and academic facilities for the American College of Radiology.

The phantom was developed by Ronald J. JaszczakJennifer Prekeges. Nuclear Medicine Instrumentation. Jones & Bartlett Publishers. 2012. p. 189. . of , and a patent was filed for in 1982. It is a cylinder containing fillable inserts that is often used with a radionuclide such as technetium-99mMattsson S, Hoeschen C. Radiation Protection in Nuclear Medicine. Springer. 2003. . p. 82. or fluorine-18.Waterstram-Rich KM, Christian PE. Nuclear Medicine and PET/CT. 7th Ed. Elsevier Health Sciences, 2013. p. 345. .

Although the phantom can be used for acceptance testing, the National Electrical Manufacturers Association recommends a 30 million count acquisition and section reconstruction of the phantom be performed quarterly.Bolus NE, Brady AB. Steves' Review of Nuclear Medicine Technology. Society of Nuclear Medicine. 4th Ed. 2011. p.177

In 1981 Ronald J. Jaszczak founded Data Spectrum Corporation, which manufactures the Jaszczak phantom and several other nuclear imaging tools, such as the Hoffman Brain phantom.


Structure and composition
Jaszczak phantoms consist of a main cylinder or tank made of acrylic plastic with several inserts. The circular phantom comes in two varieties: flanged and flangeless. The latter is recommended by the American College of Radiology for accreditation of departments. All Jaszczak phantoms have six solid spheres and six sets of 'cold' rods. In flanged models, the sizes of the spheres vary. The number of rods in each set depends on the size of the rod in that set as different models of the phantom have rods of different sizes. In flangeless models, the diameters of the spheres are 9.5, 12.7, 15.9, 19.1, 25.4 and 31.8 mm, while the rod diameters are 4.8, 6.4, 7.9, 9.5, 11.1 and 12.7 mm. Both solid spheres and rod inserts mimic cold lesions in a hot background. Spheres are used to measure the image contrast while the rods are used to investigate the image resolution in SPECT systems.Bailey DL, Humm JL, et al. Nuclear Medicine Physics: A Handbook for Teachers and Students. 2014. International Atomic Energy Agency publications. . p.563